VESSEL |
Total Capacity |
3 |
5 |
Working Maximum (L) |
2 |
3 |
MCC |
Glass & Stainless Steel |
CONTROLLER |
|
LA1 - Independent Controllers |
|
LA2 - PLC |
TEMPERATURE |
Range |
Fermentation : From
50C above coolant temperature to 500C |
Sensor |
Platinum RTD |
AGITATION |
Drive |
DC
Motor - Easily
removable - Top drive |
Range |
100-1000 rpm |
Control |
DC Drive |
Impellers |
Four / Six blade
Rushton turbine impellers. Other impellers optional |
Indication |
Digital Display |
AERATION |
Air Flow |
Standard Rotameter.
Thermal Mass Flow Meter /
Controller is offered as option |
Inlet / Exhaust
Filter |
0.2 micron
autoclavable
PTFE (Disposable) |
Condenser |
Exhaust condenser -
Glass to
minimize
evaporation |
|
Gas Mixing Unit |
Optional |
pH |
Range |
0 - 14 |
|
Probe |
Gel Filled probes |
|
Control |
Through peristaltic
pumps. |
DISSOLVED OXYGEN |
Range |
0-100% |
Probe |
Polorographic |
Control Mode |
Control through air
flow or speed or
cascade in LA2 models |
ANTIFOAM |
Control |
Level Switch and
Peristatic Pump |
ELECTRICAL DATA |
Power Basic (KVA) |
0.5 |
0.5 |
CONTINUOUS CULTURE |
|
Optional |
SOFTWARE |
|
OPC server or SCADA
package |